• Genevac
  • VirTis
  • FTS Systems
  • Hotpack

 
RS Recirculating Chiller

FTS Systems
RS Recirculating Chiller

The RS Series of recirculating chillers includes the RS33, RS33 low temperature, RS44 and RS44 low temperature systems.  These are bench top recirculating chillers with a maximum low of -10C
(-40C for low temperature systems) and a maximum high of
+35C (+75C high temperature option on RS33).
 

KEY FEATURES

  • Small compact size for bench top or under counter operation
  • Mechanically refrigerated
  • Centrifugal or positive displacement pumps
  • Wide temperature range of -40C to +75C (with low temperature and high temperature options)
  • RS232 Interface

KEY BENEFITS

  • Small footprint saving valuable lab space
  • Optimal heat removal capacity with quick ramping capability to decrease test time.
  • Meets the requirements for a large variety of applications
  • Precise temperature control of your product
  • Industrial type chiller in a compact design
  • Remote control via RS-232 interface

SPECIFICATIONS

Specifications

 

RS33
RS33LT
RS44
RS44LT
Heat Removal 
Watts/BTU
120V/60Hz  
220V/50Hz
120V/60Hz
220V/50Hz
208 or
230V/60Hz 
220V/50Hz 
208 or
230V/60Hz
220V/50Hz
   +20C
1050 / 3580
871 / 2971
350 / 1195
290 / 991
1410 / 4805
1170 / 3988
504 / 1718
418 / 1425
   +10C
880 / 3005
730 / 2494
340 / 1160
282 / 962
1230 / 4190
1020 / 3477
450 / 1534
373 / 1273
     0C
760 / 2595
630 / 2153
330 / 1125
273 / 933
1050 / 3580
871 / 2971
405 / 1381
336 / 1146
   -10C
560 / 1915
464 / 1589
285 / 970
236 / 805
-
 
370 / 1261
307 / 1046
   -15C
N/A
 
280 / 955
232 / 792
N/A
 
360 / 1227
298 / 1018
   -20C
N/A
 
275 / 938
228 / 778
N/A
 
330 / 1125
273 / 933
   -30C
N/A
 
195 / 665
161 / 551
N/A
 
285 / 971
236 / 805
   -40C
N/A
 
60 / 205
49 / 170
N/A
 
130 / 443
107 / 367
Temperature Range
-10C to +35C
(optional +75C)
-40C to +35C
(optional +75C)
-10C to +35C
-40C to +35C
Reservoir Size
2 Liters
2 Liters
2 Liters
2 Liters
Control
Digital/RS-232
Digital/RS-232
Digital/RS-232
Digital/RS-232
Stability
+/-0.1C
+/-0.1C
+/-0.1C
+/-0.1C
Heater Option
600 Watts (+75C)
600 Watts (+75C)
N/A
N/A
Pumps:    
Standard Centrifugal

2 gpm @ 4 psi

2 gpm @ 4 psi

2 gpm @ 4 psi

2 gpm @ 4 psi
Optional Positivie Displacement

1.6 gpm up to 80 psi

1.6 gpm up to 80 psi

1.6 gpm up to 80 psi

1.6 gpm up to 80 psi
Dimensions: (W x D x H)    
  Inches
13 x 21 x 19"
13 x 21 x 19"
13 x 21 x 19"
13 x 21 x 19"
  Cm
33 x 53 x 48 cm
33 x 53 x 48 cm
33 x 53 x 48 cm
33 x 53 x 48 cm
  Weight (lbs/kgs)
80/36
80/36
90/41
90/41
Electrical
120V/60hz or 220V/50Hz
120V/60hz or 220V/50Hz
208V/60Hz, 230V/60Hz or 220V/50Hz
208V/60Hz, 230V/60Hz or 220V/50Hz

APPLICATION EXAMPLES

  • Reaction vessels
  • Extractors
  • Diffusion pumps
  • Lasers
  • Sputtering systems
  • Condensers
  • Electron Microscopes
  • OEM applications

OPTIONAL COMPONENTS

To configure an RS Recirculating Chiller to fit your requirements, select options from each of the groups below.  For assistance in determining which option is right for you or to receive a quote, please contact a product specialist.  We will be glad to discuss with you the options and their importance and use.

RS-33---44-Opts---Access.JPG





Chiller-Access.JPG









LITERATURE & ADDITIONAL SPECIFICATIONS

 

Product gallery

Send to a Colleague

Your name:  
Colleague's Email:
Your message (click here):
 



Free Webinar
Key Considerations for the Successful Lyophilization of Proteins
Presented by Jeff Blue, Merck

Click For More Info



SP Scientific is proud to be a sponsor of
Freeze Drying of Pharmaceuticals and Biologicals Conference
9/28 - 10/1
Garmisch Germany
Link to Information



LYOLEARN WEBINAR ARCHIVE
Freeze drying topics
Given by freeze drying experts

Recorded for you to listen to on demand
FOLLOW THIS LINK